What Is the Bradford Assay?
The Bradford assay is a protein quantification method introduced by Marion M. Bradford in 1976. It relies on the interaction between proteins and the Coomassie Brilliant Blue G-250 dye. When the dye binds to protein, its absorption characteristics change, producing a measurable increase in absorbance around 595 nm.
Because of its speed and straightforward workflow, the Bradford assay is widely used in molecular biology, biochemistry, protein purification, cell biology, and other research applications where protein concentration needs to be estimated before downstream experiments.
Bradford Assay Protocol: Principle
In a typical Bradford assay protocol, a series of protein standards with known concentrations is prepared first. The standards and unknown samples are then reacted with Bradford reagent. After incubation, absorbance is measured at 595 nm. The measured absorbance values of the standards are used to generate a calibration curve, which is then used to determine the protein concentration of unknown samples by interpolation.
Bradford Assay Protocol: Materials
| Bradford reagent | Coomassie Brilliant Blue G-250-based reagent |
| Protein standard | BSA or another suitable protein standard |
| Unknown samples | Protein-containing research samples |
| Microplate or tubes | Depending on the selected assay format |
| Reader | Microplate reader or spectrophotometer capable of measuring 595 nm |
Step-by-Step Bradford Assay Protocol
1. Prepare the Protein Standards
Prepare a series of protein standards covering the expected concentration range of the samples. Bovine serum albumin (BSA) is commonly used as a standard for Bradford protein assays. Bio-Rad recommends preparing multiple concentrations covering the linear range of the selected assay format.
2. Prepare the Samples
Dilute unknown samples when necessary so that their protein concentrations fall within the useful range of the assay. Standards and unknown samples should be prepared in compatible conditions and processed in the same way to improve comparability.
3. Add Bradford Reagent
Add the appropriate amount of Bradford reagent to each standard and sample according to the specific assay format and manufacturer's instructions. Mix thoroughly to ensure consistent interaction between the protein and dye.
4. Incubate
Allow the reaction to develop at room temperature for the period specified by the assay protocol. Common Bradford procedures use an incubation period of approximately 5 to 60 minutes, depending on the assay format and reagent system.
5. Measure Absorbance at 595 nm
Measure the absorbance of standards and samples at 595 nm using a suitable spectrophotometer or microplate reader. The dye-protein interaction produces the measurable color response used for quantification.
6. Generate the Standard Curve
Plot the absorbance values of the known standards against their protein concentrations. The resulting standard curve is used to calculate the concentration of the unknown samples by interpolation. Running a new standard curve with each assay is recommended.
7. Calculate Protein Concentration
Determine the concentration of each unknown from the standard curve. If a sample was diluted before the assay, multiply the calculated concentration by the appropriate dilution factor to obtain the concentration of the original sample.
Bradford Assay Protocol in a 96-Well Plate
The microplate format is particularly useful when many samples must be analyzed. A typical Bradford microplate workflow involves dispensing standards and samples into the wells, adding Bradford reagent, mixing, incubating at room temperature, and measuring absorbance at 595 nm. Exact volumes depend on the kit and assay format.
For example, Bio-Rad's published microplate standard assay uses small sample volumes followed by Bradford dye reagent, incubation at room temperature, and absorbance measurement at 595 nm. The manufacturer recommends processing standards and samples consistently and running replicates.
Example Bradford Assay Workflow
| Step | Action |
| 1 | Prepare protein standards and unknown samples. |
| 2 | Add standards and samples to tubes or microplate wells. |
| 3 | Add Bradford reagent according to the selected assay procedure. |
| 4 | Mix and allow color to develop. |
| 5 | Measure absorbance at 595 nm. |
| 6 | Construct the standard curve. |
| 7 | Determine unknown protein concentrations. |
How to Make a Bradford Standard Curve
The standard curve is one of the most important parts of the Bradford assay protocol. Prepare several known protein concentrations and measure their absorbance under identical conditions. Plot protein concentration on the x-axis and absorbance at 595 nm on the y-axis.
The appropriate mathematical model depends on the behavior of the standards over the selected concentration range. Samples should ideally fall within the validated working range rather than being extrapolated far beyond the standards.
Bradford Assay Tips for Better Results
Run a standard curve with each assay.
Variations in reagent, handling and instrument conditions can affect measurements, so a fresh calibration curve improves consistency.
Use replicates.
Running standards and samples in replicate can help identify pipetting errors and improve confidence in the calculated concentrations.
Keep standards and samples under comparable conditions.
Processing the standards and unknowns in the same way helps ensure that differences in absorbance reflect differences in protein concentration.
Watch for interfering substances.
Although Bradford assays are compatible with many compounds found in protein preparations, some buffers and substances can interfere with the assay. Dilution or an alternative protein assay may be necessary in problematic sample matrices.
Bradford Assay Protocol: Common Errors
Several factors can reduce the reliability of Bradford measurements. Common issues include using an inappropriate standard concentration range, inconsistent incubation times, inaccurate pipetting, insufficient mixing, and measuring samples outside the useful response range.
Another important consideration is protein-to-protein variation. Different proteins can produce different color responses with Coomassie dye, so the calculated concentration should be interpreted as a relative protein concentration unless the standard closely represents the protein being measured.
Bradford Assay Applications
The Bradford method can be incorporated into a wide range of research workflows, including:
Protein purification | Cell lysate analysis | Western blot sample preparation | Enzyme studies | Biochemistry | Molecular Biology
Bradford Assay Kit for Protein Quantification
A ready-to-use Bradford assay kit can simplify routine protein quantification by providing the reagents needed to perform the colorimetric assay. Gentaur's Bradford Assay Kit is listed in a 1,000-test format and can be used as part of laboratory protein quantification workflows.
Bradford Assay Protocol FAQs
What wavelength is used in the Bradford assay?
Bradford assays are commonly measured at 595 nm, where the protein-bound Coomassie dye produces the characteristic colorimetric response.
What protein standard is commonly used?
Bovine serum albumin (BSA) is one of the most commonly used standards for Bradford protein assays. Bovine gamma-globulin is another standard used in some assay systems.
How is protein concentration calculated?
Protein concentration is determined by comparing the sample's absorbance with a standard curve generated from proteins of known concentration. The calculated value should then be adjusted for any sample dilution.
Can the Bradford assay be used with any protein sample?
The assay can be used for many protein-containing samples, but the response can vary between proteins and some buffer components or other substances can interfere. Sample compatibility should therefore be evaluated before quantitative analysis.
References
Bradford MM. A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye binding. Analytical Biochemistry. 1976;72:248-254. DOI: 10.1016/0003-2697(76)90527-3.
Bio-Rad Laboratories. Bradford Protein Assay and Protein Assay technical guidance, including standard curve generation, assay formats, 595 nm measurement and recommended practices.
Research Use Only: The information provided in this article is intended for laboratory research and educational purposes. Always follow the specific instructions supplied with the Bradford assay kit being used.